Skip to main content

Pressure Gauge Calibration Procedure

 A Pressure Gauge is a fluid intensity measurement device. The pressure gauge is required for the setup & tuning of fluid power machines and is indispensable in troubleshooting them without pressure gauges.

The fluid power system would be both unpredictable & unreliable. Gauges help us to ensure there are no leaks or pressure changes that could affect the operating condition of the processing system in industries.

Pressure Gauge Class

There are two sets of standards that define the accuracy grade of pressure gauges.

ASME B40.1 defines the accuracy grades of dial-type gauges while ASME B40.7 defines the accuracy grades of digital gauges.

ASME B40.1


Why we are discussing the above standards? because we should know all the basics & Standards before performing the calibration of Pressure Gauge. Then only we will be able to identify whether the Pressure Gauge under calibration is acceptable or not.

Master Equipment’s required for performing Calibration:
1. Pressure Comparator
A Pressure Comparator operates in a similar manner to a dead-weight tester for the generation of pressure until the desired pressure is achieved.

In this case, we compare the UUC (Unit Under Calibration) with the Master Gauge.
2. Master Gauge
It is a digital gauge that is used to calibrate measurement instruments in ISO 9000 qualified laboratories. This calibration equipment can offer an accuracy of ±0.1%, mainly depends on the manufacturer.


Calibration Procedure
 

1. As shown in Picture below, first identify the basics parameters on UUC (Unit Under Calibration)

a) Class of Pressure Gauge

b) Minimum/Maximum Range of Pressure gauge

c) Least Count

d) Measuring Unit
Note: for identifying the Least Count, Calculate the lines in between major line, in the above picture the

2 bar is the first reading but followed by 10 small lines i.e.

2/10 = 0.2 which means 0.2 is the least count of this Pressure gauge.

2. According to the class of pressure gauge, we will able to identify the pressure gauge acceptable limits.

3. Release the pressure of pressure comparator before using it by the movable wheel.

4. Place the Master Gauge on the Left-Hand side as shown in the setup figure.
5. Select the unit/least count as per UUC on Master Gauge with selection keys.

6. Place the UUC (pressure gauges to be calibrated) to Right Hand Side, as shown in the figure.

7. Use the handwheel to pump the liquid in the comparator for proceeding calibration steps until it reaches the desired range.

8. Example:

Suppose the UUC Gauge shown in the above picture has following parameters,

Class: 1.0

Min Range: 0 Bar

Max Range: 10 Bar

Least Count: 0.2 Bar

Unit: Bar

Here we will calibrate the pressure gauge for 2, 4, 6, 8, 10 bar readings in comparison of Master Gauge.

We need to pump the oil till we reach 2 bar pressure on Master Gauge and note down the readings of Master as well as the UUC at that point.

9. Next, rotate the handwheel to increase the pressure to 4 bar, note down the readings of master and UUC gauges.

10. Similarly, take the readings for 6bar, 8bar, and 10bar.

9. We have taken the readings from 0 to 10 bar in an upward direction. Now do the same in the downward direction, meaning that decrease the pressure from 10 bar to 0 bar and note down the readings of both the gauges. We have to take the readings for 10bar, 8bar, 6bar, 4bar, 2bar, and 0bar.

Also Read: Why Upward & Downward Calibration?

10. After completing the above steps, release the pressure of the Comparator and remove the UUC and Master Gauge.

Pressure Gauge Calibration Certificate
For your reference the Final Calibration certificate will be followed by;







Comments

Popular posts from this blog

Ferrules and Cross Ferruling

 Ferrules are identification labels provided for every wire terminations in an instrument, equipment, or electrical/instrumentation control panels. These tube-shaped sleeves can be inserted easily on each individual wire in a multi-core cable. In earlier days fixed digits/letters are used as ferrules, but now Instrumentation engineers/technicians prints out desired ferrules by using a ferrule printing machine. Typical Ferrule The numbers/ letters on the ferrules will be given as per the approved electrical hook up or loop diagrams. This helps technicians to easily identify a particular loop/wiring from a series of terminal blocks and to troubleshoot the desired terminal connection. Separate numbers on the ferrules distinguish the positive and negative polarities of wires, thus ensure the polarity protection of the instrument. Cross Ferruling  As a wire is connected on its both ends, it is quite useful to use a cross reference method for wire identification. Unlike normal ferru...

What is a Torbar? – Averaging Pitot Tubes

 The Torbar is employed for flow measurement of liquids, gases, or steam in circular, square, or rectangular section ducts for large flow rates. The Torbar is an insertion type multi-port self-averaging primary sensor for flow measurement. Torbar TORBAR is a set of Pitot tubes mounted on a bar across the pipeline with no moving parts. An averaging Pitot tube is a technology, while TORBAR is a manufacturing brand name. There are several brands available in the market with VERABAR, ANNUBAR, etc. Averaging Pitot Tube Principle Purpose Averaging Pitot tube can be employed when the average velocity of the flow profile, rather than the velocity in a specific point in the cross-section is desired. Averaging Pitot Tubes Principle It measures the differential pressure between the static pressure tap and the tap of full pressure of a stream. Thus such magnitude of differential pressure is directly proportional to the square of the flow rate. Working The TORBAR is designed in such a way that ...

Dissolved Oxygen Analyzer Working Principle

 What is dissolved oxygen ? Dissolved oxygen refers to oxygen dissolved in water. Its concentration is expressed as the amount of oxygen per unit volume and the unit is mg/L. Biologically, oxygen is an essential element for respiration of underwater life and also acts as a chemical oxidizer. The solubility of oxygen in water is affected by water temperature, salinity, barometric pressure, etc. and decreases as water temperature rises. Measurement of dissolved oxygen by the membrane electrode method The membrane electrode method measures a diffusion current or reduction current generated by the concentration of dissolved oxygen or partial pressure of oxygen to obtain the concentration of dissolved oxygen. This method is not affected by the pH value of water being measured, oxidation and reduction substances, color, turbidity, etc. and the measurement method offers good reproducibility. When a sensor is inserted into water, an air layer forms on the membrane (Teflon membrane). The ox...