As-found and As-left Documentation during Calibration

 As-found and as-left Documentation during Calibration

Calibration is the comparison of a measurement device (an unknown) against an equal or better standard. A standard in a measurement is considered the reference; it is the one in the comparison taken to be the more correct of the two. Calibration finds out how far the unknown is from the standard.

A “typical” commercial calibration uses the manufacturer’s calibration procedure and is performed with a reference standard at least four times more accurate than the instrument under test.

An important principle in calibration practice is to document every instrument’s calibration as it was found and as it was left after adjustments were made. The purpose for documenting both conditions is to make data available for calculating instrument drift over time. If only one of these conditions is documented during each calibration event, it will be difficult to determine how well an instrument is holding its calibration over long periods of time.

Excessive drift is often an indicator of impending failure, which is vital for any program of predictive maintenance or quality control. Typically, the format for documenting both As-Found and As-Left data is a simple table showing the points of calibration, the ideal instrument responses, the actual instrument responses, and the calculated error at each point.

The following table is an example for a pressure transmitter with a range of 0 to 200 PSI over a five-point scale:

The following photograph shows a single-point “As-Found” calibration report on a temperature indicating controller, showing the temperature of the calibration standard (−78.112 degrees Celsius), the display of the instrument under test (IUT, −79 degrees Celsius), and the error between the two (−0.888 degrees Celsius):
Note that the mathematical sign of the error is important. An instrument that registers −79 degrees when it should register −78.112 degrees exhibits a negative error, since its response is lower (i.e. more negative) than it should be.

Expressed mathematically: Error = IUT − Standard. When the error must be expressed in percentage of span, the formula becomes:

Calibration Terms
As found data : The reading of the instrument before it is adjusted.

As left data : The reading of the instrument after adjustment or “same as found,” if no adjustment was made.

Optimization : Adjusting a measuring instrument to make it more accurate is NOT part of a typical calibration and is frequently referred to as “optimizing” or “nominalizing” an instrument.

Out-of-tolerance (OOT) condition : When an instrument’s performance is outside its specifications, it is considered an outof- tolerance (OOT) condition, resulting in the need to adjust the instrument back into specification.

Limited calibration : It may be more cost effective to have a limited calibration when only certain functions of an instrument are not utilized by the user.

Test uncertainty ratio (TUR) : This is the ratio of the accuracy of the instrument under test compared to the accuracy of the reference standard.

Verification : Typically referring to a functional check or performing a comparison with an instrument not classified as a reference standard for the full measurement.

Measurement Uncertainty : Influences in the measurement causing uncertainty (%). Examples: temp., pressure., humidity, fittings etc.

MPE – Maximum Permissible Error  : Tolerance allowed in process by the customer for determining whether or not an instrument passes or fails (instrument & process tolerance)

Example Calibration Record Sheet :






Comments

Popular posts from this blog

Ferrules and Cross Ferruling

 Ferrules are identification labels provided for every wire terminations in an instrument, equipment, or electrical/instrumentation control panels. These tube-shaped sleeves can be inserted easily on each individual wire in a multi-core cable. In earlier days fixed digits/letters are used as ferrules, but now Instrumentation engineers/technicians prints out desired ferrules by using a ferrule printing machine. Typical Ferrule The numbers/ letters on the ferrules will be given as per the approved electrical hook up or loop diagrams. This helps technicians to easily identify a particular loop/wiring from a series of terminal blocks and to troubleshoot the desired terminal connection. Separate numbers on the ferrules distinguish the positive and negative polarities of wires, thus ensure the polarity protection of the instrument. Cross Ferruling  As a wire is connected on its both ends, it is quite useful to use a cross reference method for wire identification. Unlike normal ferru...

What is a Torbar? – Averaging Pitot Tubes

 The Torbar is employed for flow measurement of liquids, gases, or steam in circular, square, or rectangular section ducts for large flow rates. The Torbar is an insertion type multi-port self-averaging primary sensor for flow measurement. Torbar TORBAR is a set of Pitot tubes mounted on a bar across the pipeline with no moving parts. An averaging Pitot tube is a technology, while TORBAR is a manufacturing brand name. There are several brands available in the market with VERABAR, ANNUBAR, etc. Averaging Pitot Tube Principle Purpose Averaging Pitot tube can be employed when the average velocity of the flow profile, rather than the velocity in a specific point in the cross-section is desired. Averaging Pitot Tubes Principle It measures the differential pressure between the static pressure tap and the tap of full pressure of a stream. Thus such magnitude of differential pressure is directly proportional to the square of the flow rate. Working The TORBAR is designed in such a way that ...

Thermowell Insertion and Immersion Length

 The thermowell is the industrial housing for temperature elements like RTD or thermocouple or temperature gauges. Thermowell is directly exposed to process fluid/media. Inside the thermowell, element/gauge is installed. Thermowell transfers the heat to the temperature element. Depending upon the element type, the fluid temperature can be known. So Thermowell protects the temperature element against harsh conditions and process fluid. It enables removal of the temperature element for replacement, repair, or testing without affecting the process system. Generally, thermowells are provided with all-temperature Elements like RTD or Thermocouple or temperature gauge. A few exceptions would be bearing RTD, skin type thermocouple, air duct thermometer, etc where thermowell is not applicable / required. When the thermowell assembly is inserted into the pipe, it is subjected to vibration. This vibration is caused by vortices to be generated alternately on either side of the thermowell. The...