Calibration and Troubleshooting of Oxygen Analyzer

 Calibration of Oxygen Analyzer

Calibration of most types of oxygen analyzers can be performed manually or by an automatic calibration option. The analyzer controller can be programmed to perform the automatic calibration at specific time intervals, review the calibration results and if they are within programmable parameters, update the controller with the new calibration data.

The procedures are the same for manual or automatic calibrations with the exception that in manual mode, the calibration data has to be manually accepted after the calibration runs.

The calibration procedure is very similar to each type of analyzer, therefore, it is recommended to following the manufacturer’s procedure. The calibration procedures are typically programmer orientated and involve two calibrations gases with the results adjusting the slope of the calibration line within the software.

To Calibrate

*Zero and span gases are connected to the analyzer.

*With the zero-gas flowing through the analyzer, allow the readings to stabilize, then correct the reading to equal the zero gas concentration.

*With the span-gas flowing through the analyzer, allow the readings to stabilize, then correct the reading to equal the span gas concentration.


Troubleshooting of Oxygen Analyzer
Faults with the oxygen analyzer output can be generated from several courses:

Sampling System

The sampling system should be periodically checked for proper flows and pressures and filtration and moisture removal.

Low readings are an indication of improper sampling.

Sensor

Zirconia Cell

The inaccurate readings could result from:


Heater temperature thermocouples

Heater itself

Heater temperature controller

Contaminated electrodes

Contaminated zirconia cell

All items should be checked according to the manufacturer’s procedures and diagnostics


Thermoparamagnetic

The inaccurate readings could result from:


Thermistors

Electronic sensing circuit

Auto-zero function

Changing magnetic field

All items should be checked according to the manufacturer’s procedures and diagnostics


Glass Sphere Paramagnetic

The inaccurate readings could result from:


Electronic sensing circuit

Light emitter

Photodiodes

Changing magnetic field

All items should be checked according to the manufacturer’s procedures and diagnostics


Electronics

All analyzers are equipped with extensive self-diagnostics and troubleshooting guides by the manufacturer, which should be followed for any alarms or inaccurate readings.


Author: Mohammed Khaleel

Comments

Popular posts from this blog

What is a Torbar? – Averaging Pitot Tubes

 The Torbar is employed for flow measurement of liquids, gases, or steam in circular, square, or rectangular section ducts for large flow rates. The Torbar is an insertion type multi-port self-averaging primary sensor for flow measurement. Torbar TORBAR is a set of Pitot tubes mounted on a bar across the pipeline with no moving parts. An averaging Pitot tube is a technology, while TORBAR is a manufacturing brand name. There are several brands available in the market with VERABAR, ANNUBAR, etc. Averaging Pitot Tube Principle Purpose Averaging Pitot tube can be employed when the average velocity of the flow profile, rather than the velocity in a specific point in the cross-section is desired. Averaging Pitot Tubes Principle It measures the differential pressure between the static pressure tap and the tap of full pressure of a stream. Thus such magnitude of differential pressure is directly proportional to the square of the flow rate. Working The TORBAR is designed in such a way that ...

Ferrules and Cross Ferruling

 Ferrules are identification labels provided for every wire terminations in an instrument, equipment, or electrical/instrumentation control panels. These tube-shaped sleeves can be inserted easily on each individual wire in a multi-core cable. In earlier days fixed digits/letters are used as ferrules, but now Instrumentation engineers/technicians prints out desired ferrules by using a ferrule printing machine. Typical Ferrule The numbers/ letters on the ferrules will be given as per the approved electrical hook up or loop diagrams. This helps technicians to easily identify a particular loop/wiring from a series of terminal blocks and to troubleshoot the desired terminal connection. Separate numbers on the ferrules distinguish the positive and negative polarities of wires, thus ensure the polarity protection of the instrument. Cross Ferruling  As a wire is connected on its both ends, it is quite useful to use a cross reference method for wire identification. Unlike normal ferru...

Thermowell Insertion and Immersion Length

 The thermowell is the industrial housing for temperature elements like RTD or thermocouple or temperature gauges. Thermowell is directly exposed to process fluid/media. Inside the thermowell, element/gauge is installed. Thermowell transfers the heat to the temperature element. Depending upon the element type, the fluid temperature can be known. So Thermowell protects the temperature element against harsh conditions and process fluid. It enables removal of the temperature element for replacement, repair, or testing without affecting the process system. Generally, thermowells are provided with all-temperature Elements like RTD or Thermocouple or temperature gauge. A few exceptions would be bearing RTD, skin type thermocouple, air duct thermometer, etc where thermowell is not applicable / required. When the thermowell assembly is inserted into the pipe, it is subjected to vibration. This vibration is caused by vortices to be generated alternately on either side of the thermowell. The...