Ambient Air Quality Monitoring System Principle

 AAQMS (ambient air quality monitoring System) for continuous monitoring of pollutant gases like NO2, SO2, O3, CO and suspended particulate matter like PM10 and PM2.5

Ambient Air Quality Monitoring Parameters As per norms below mentioned parameters are mandatory for air quality.

CO : Carbon Monoxide

SO2: Sulphur Dioxide

NO2 : Nitrogen Dioxide

O3: Ozone

PM10: Particulate Matter

The following provides information about the basic principles used for measurement.

Sulfur dioxide (SO2) :

Sulfur dioxide (SO2) is measured continuously by instruments using the fundamental principle of fluorescence. Fluorescent analyzers irradiate an ambient air sample with ultraviolet light at 213.8 nm. Sulfur dioxide gas molecules absorb a portion of this energy, then re-emit the energy at a characteristic wavelength of light. This light energy emitted by SO2 molecules is sensed by a photomultiplier tube and converted to an electronic signal proportional to the concentration of SO2present.

Nitrogen dioxide (NO2)  :

Continuous monitoring of nitrogen dioxide (NO2) is based on a chemiluminescent reaction between nitric oxide (NO) and ozone (O3). When these two gases react, light energy at a specific wavelength is produced. In the monitor, ambient air is drawn along two paths. In the first path, the air is reacted directly with ozone, and the light energy produced is proportional to the amount of NO in the air. In the second path, the air is reacted with ozone after it passes through a catalytic reduction surface. The reduction surface converts NO2 to NO and the light energy produced is a measure of the total oxides of nitrogen in the air sample. The electronic difference of these two signals yields the concentration of NO2. All concentrations for NO2 are given in parts per million (ppm).

Carbon monoxide (CO) is monitored continuously by analyzers that operate on the infrared absorption principle. Ambient air is drawn into a sample chamber and a beam of infrared light is passed through it. CO absorbs infrared radiation, and any decrease in the intensity of the beam is due to the presence of CO molecules. This decrease is directly related to the concentration of CO in the air. A special detector measures the difference in the radiation between this beam and a duplicate beam passing through a reference chamber with no CO present. This difference in intensity is electronically translated into a reading of the CO present in the ambient air, measured in parts per million (ppm).

Ozone :

Ozone is measured continuously by analyzers that operate using the monochromatic ultraviolet absorption spectrophotometry principle. As ultraviolet light at 253.7 nm is passed through the optic bench, a fixed quantity of “zero air” and ambient air are drawn into the bench. The intensity of the ultraviolet radiation traversing the optics bench is attenuated by the ozone present in the ambient sample. This attenuated signal is detected and compared with the unattenuated signal from the “zero air” cycle. This difference in intensity is electronically translated into a reading of ozone present in the ambient air.

Lead (Pb) :

Lead (Pb) concentrations in ambient air are determined by the reference method promulgated by the U.S. EPA. The Pb sample is collected on a filter using a high-volume air sampler and the TSP method. In this method, two 3/4″ X 8″ portions of the teaspoon filters are washed with hot, diluted nitric acid. The Pb compounds are dissolved into the acid solutions. The solution is then analyzed by the atomic absorption technique to determine the amount of lead.

Particulate Matter :

The shorthand PM10 is defined as particulate matter with an aerodynamic diameter less than or equal to a nominal 10 micrometers. The reference method for the measurement of atmospheric particulate matter as PM10 is based on selection of PM10 particles by internal separation, followed by filtration and gravimetric determination of the PM10 mass on a filter substrate. Selection of PM10 is accomplished by a size selective head that is symmetrical designed and contains nine circular acceleration nozzles. By virtue of their larger momentum, particles greater than 10 micron aerodynamic diameter impact onto a greased impaction shim. The PM10 particles smaller than 10 microns are carried vertically upward by the air flow and down 16 vent tubes to a 8-by-10 inch quartz-fiber filter, where they are collected.


Source : epa.ohio

Comments

Popular posts from this blog

Ferrules and Cross Ferruling

 Ferrules are identification labels provided for every wire terminations in an instrument, equipment, or electrical/instrumentation control panels. These tube-shaped sleeves can be inserted easily on each individual wire in a multi-core cable. In earlier days fixed digits/letters are used as ferrules, but now Instrumentation engineers/technicians prints out desired ferrules by using a ferrule printing machine. Typical Ferrule The numbers/ letters on the ferrules will be given as per the approved electrical hook up or loop diagrams. This helps technicians to easily identify a particular loop/wiring from a series of terminal blocks and to troubleshoot the desired terminal connection. Separate numbers on the ferrules distinguish the positive and negative polarities of wires, thus ensure the polarity protection of the instrument. Cross Ferruling  As a wire is connected on its both ends, it is quite useful to use a cross reference method for wire identification. Unlike normal ferru...

What is a Torbar? – Averaging Pitot Tubes

 The Torbar is employed for flow measurement of liquids, gases, or steam in circular, square, or rectangular section ducts for large flow rates. The Torbar is an insertion type multi-port self-averaging primary sensor for flow measurement. Torbar TORBAR is a set of Pitot tubes mounted on a bar across the pipeline with no moving parts. An averaging Pitot tube is a technology, while TORBAR is a manufacturing brand name. There are several brands available in the market with VERABAR, ANNUBAR, etc. Averaging Pitot Tube Principle Purpose Averaging Pitot tube can be employed when the average velocity of the flow profile, rather than the velocity in a specific point in the cross-section is desired. Averaging Pitot Tubes Principle It measures the differential pressure between the static pressure tap and the tap of full pressure of a stream. Thus such magnitude of differential pressure is directly proportional to the square of the flow rate. Working The TORBAR is designed in such a way that ...

Dissolved Oxygen Analyzer Working Principle

 What is dissolved oxygen ? Dissolved oxygen refers to oxygen dissolved in water. Its concentration is expressed as the amount of oxygen per unit volume and the unit is mg/L. Biologically, oxygen is an essential element for respiration of underwater life and also acts as a chemical oxidizer. The solubility of oxygen in water is affected by water temperature, salinity, barometric pressure, etc. and decreases as water temperature rises. Measurement of dissolved oxygen by the membrane electrode method The membrane electrode method measures a diffusion current or reduction current generated by the concentration of dissolved oxygen or partial pressure of oxygen to obtain the concentration of dissolved oxygen. This method is not affected by the pH value of water being measured, oxidation and reduction substances, color, turbidity, etc. and the measurement method offers good reproducibility. When a sensor is inserted into water, an air layer forms on the membrane (Teflon membrane). The ox...